Dimensions / Weight
| W | H | D | |
|---|---|---|---|
| Dimensions | 212.8 mm (8.38 inch) | 88.3 mm (3.48 inch) | 272.3 mm (10.72 inch) |
Digital Multimeter
DME1700 Series
Price range: 158,000円 through 198,000円
- Resolution : 6 1/2 digit
- Enhanced visibility: 4.3-inch TFT color LCD
- Basic measurement function
- DC voltage: 100 mV, 1 V, 10 V, 100 V, 1000 V
- AC voltage: 100 mV, 1 V, 10 V, 100 V, 750 V
- DC current: 10 mA, 100 mA, 1 A, 3 A
- AC current: 10 mA, 100 mA, 1 A, 3 A
- 2-wire/4-wire resistance: 100 Ω, 1 kΩ, 10 kΩ, 100 kΩ, 1 MΩ, 10 MΩ, 100 MΩ (100 MΩ range is 2-wire resistance only)
- Frequency and period measurement: 3 Hz to 300 kHz
- Other measurement functions: Continuity test, diode test, temperature measurement
- Measurement range: Selectable in auto range or manual range
- Maximum sample count: 1,500,000 (DME1701 only)
- Standard communication interfaces included (DME1701)
LAN, USB *Options: GPIB, RS232C (selected at time of order)
Line-up
| Image | SKU | Interface | Maximum reading speed | 標準価格 | WEBかんたんお見積り |
|---|---|---|---|---|---|
![]() | DME1700 | – | 300rdg/s | 158,000円 (税込:173,800円) | |
![]() | DME1701 | USB, LAN | 10000rdg/s | 198,000円 (税込:217,800円) |
Product Overview
Precise measurements and quick readings for various testing requirements.
Measurement data can be exported directly to a USB flash drive as a CSV file.
The DME1700 Series digital multimeters have 6.‑digit resolution. This series provides various measurement functions, including voltage, current, resistance, frequency, and temperature, making it suitable for many evaluation and measurement applications.
This series features a 4.3‑inch TFT color LCD. Auto-ranging provides user‑friendly operation. Moreover, DME1700 achieves an accuracy of 0.0035% for DC voltage (10 V range) and 0.01% for resistance (10 kΩ range)*. It delivers dependable measurements in environments that require high precision.
*Value for the DME1701 within one year of calibration.
At the fastest integration time (PLC), the DME1700 achieves a maximum of 300 rdg/s (0.2 PLC) and the DME1701 a maximum of 10,000 rdg/s (0.006 PLC), which is ideal for automated measurement systems that require fast, high-volume measurements.
DME1701 comes standard with LAN. USB, GPIB* and RS-232C* interfaces are available as options, providing flexibility for embedded applications. When used with an optional scanner card, the unit can measure up to 20 channels, making it well-suited for multi-point evaluation.
*Selected at time of order.
Specifications
DC voltage measurement (DCV)
| Item | DME1700 | DME1701 | ||||
|---|---|---|---|---|---|---|
| Measurement range | Resolution | Input resistance | Accuracy (1 year) | Temperature coefficient/°C | Accuracy (1 year) | Temperature coefficient/°C |
| 100mV | 0.1μV | >10GΩ | 0.0090+0.0065 | 0.0005+0.0005 | 0.0050+0.0035 | 0.0005+0.0005 |
| 1 V | 1.0μV | 0.0080+0.0010 | 0.0005+0.0001 | 0.0040+0.0007 | 0.0005+0.0001 | |
| 10V | 10μV | 0.0075+0.0005 | 0.0005+0.0001 | 0.0035+0.0005 | 0.0005+0.0001 | |
| 100V | 100μV | 10MΩ | 0.0085+0.0006 | 0.0010+0.0001 | 0.0060+0.0006 | 0.0010+0.0001 |
| 1000V | 1 mV | 0.0085+0.0010 | 0.0010+0.0001 | 0.0060+0.0010 | 0.0010+0.0001 | |
DC current measurement (DCI)
| Item | DME1700 | DME1701 | ||||
|---|---|---|---|---|---|---|
| Measurement range | Resolution | Shunt resistance | Accuracy (1 year) | Temperature coefficient/°C | Accuracy (1 year) | Temperature coefficient/°C |
| 10mA | 10nA | 5.1Ω | 0.050+0.020 | 0.0020+0.0020 | 0.050+0.020 | 0.0020+0.0020 |
| 100mA | 100nA | 0.050+0.005 | 0.0020+0.0005 | 0.050+0.005 | 0.0020+0.0005 | |
| 1A | 1μA | 0.1Ω | 0.100+0.010 | 0.0050+0.0010 | 0.100+0.010 | 0.0050+0.0010 |
| three a (baseball) | 10μA | 0.200+0.002 | 0.0050+0.0020 | 0.200+0.020 | 0.0050+0.0020 | |
2-wire/4-wire resistance measurement (Ω2W/Ω4W)
| Item | DME1700 | DME1701 | ||||
|---|---|---|---|---|---|---|
| Measurement range | Resolution | Test current | Accuracy (1 year) | Temperature coefficient/°C | Accuracy (1 year) | Temperature coefficient/°C |
| 100Ω | 100μΩ | 1 mA | 0.014+0.007 | 0.0006+0.0005 | 0.010+0.004 | 0.0006+0.0005 |
| 1kΩ | 1MΩ | 0.014+0.001 | 0.0006+0.0001 | 0.010+0.001 | 0.0006+0.0001 | |
| 10kΩ | 10mΩ | 100μA | 0.014+0.001 | 0.0006+0.0001 | 0.010+0.001 | 0.0006+0.0001 |
| 100kΩ | 100mΩ | 10μA | 0.014+0.001 | 0.0006+0.0001 | 0.010+0.001 | 0.0006+0.0001 |
| 1MΩ | 1Ω | 0.014+0.001 | 0.0010+0.0002 | 0.010+0.001 | 0.0010+0.0002 | |
| 10MΩ | 10Ω | 1μA | 0.040+0.001 | 0.0030+0.0004 | 0.040+0.001 | 0.0030+0.0004 |
| ≧100MΩ*1 | 100Ω | 1μA | 10MΩ | 0.800+0.010 | 0.1500+0.0002 | 0.800+0.010 | 0.1500+0.0002 |
*1 The ≥ 100 MΩ range is only available for 2-wire resistance measurement (Ω2W).
Diode test
| Item | DME1700 | DME1701 | ||||
|---|---|---|---|---|---|---|
| Measurement range | Resolution | Test current (±5 %) | Accuracy (1 year) | Temperature coefficient/°C | Accuracy (1 year) | Temperature coefficient/°C |
| 10V (fixed) | 10μV | 1 mA | 0.01+0.03 | 0.001+0.002 | 0.01+0.03 | 0.001+0.002 |
Continuity test
| Item | DME1700 | DME1701 | ||||
|---|---|---|---|---|---|---|
| Measurement range | Resolution | Test current (±5 %) | Accuracy (1 year) | Temperature coefficient/°C | Accuracy (1 year) | Temperature coefficient/°C |
| 1kΩ (fixed) | 10mΩ | 1 mA | 0.01+0.03 | 0.001+0.002 | 0.01+0.03 | 0.001+0.002 |
DC measurement characteristics
| Item | DME1700 | DME1701 | ||
|---|---|---|---|---|
| DC voltage measurement (DCV) | Over-range | For 100 mV to 100 V ranges | 20 % over-range allowed | |
| For 1000 V range | 1 % over-range allowed | |||
| A/D converter linearity | For 10 V range | 0.0001 % of measurement range | ||
| Input impedance selection *1 | For 100 mV to 10 V ranges | 10 MΩ ± 1 % or >>10 GΩ selectable | ||
| For 100 V to 1000 V range | 10 MΩ ± 1 % only | |||
| Input bias current | At 23 °C (73.4 °F) | Less than 50 pA | ||
| Common mode voltage (Between LO terminal and chassis) | Max. 500 Vpk | |||
| Auto zero adjustment | Within ± 1 °C (± 1.8 °F) and within 10 min of the last auto zero adjustment | Adding ± (0.0002 % of measurement range + 3 μV) to accuracy | ||
| additional error for disabled setting | Within ± 5 °C (± 9 °F) and within 60 min after the last auto zero adjustment | Adding ± (0.001 % of measurement range + 10 μV) to accuracy | ||
| DC current measurement (DCI) | Over-range | For 10 mA to 1 A ranges | 20 % over-range allowed | |
| For 3 A range | No over-range | |||
| Common mode current | 50 Hz or 60 Hz | < 600 nApk-pk | ||
*1 Specifications for parallel connection of less than 400 pF.
Frequency and period measurement (FREQ/PERIOD)
| Input signal range | Frequency measurement range | DME1700 | DME1701 | ||
|---|---|---|---|---|---|
| Accuracy (1 year) | Temperature coefficient/°C | Accuracy (1 year) | Temperature coefficient/°C | ||
| 100 mV to 750 V (10 mA to 3 A) | 3 Hz to 10 Hz | 0.1 | 0.005 | 0.1 | 0.1 |
| 10 Hz to 100 Hz | 0.05 | 0.005 | 0.03 | 0.035 | |
| 100 Hz to 1 kHz | 0.02 | 0.001 | 0.01 | 0.015 | |
| 1 kHz to 300 kHz | 0.012 | 0.001 | 0.01 | 0.015 | |
Measurement characteristics (over-range)
Measurement characteristics (over-range)
| Item | DME1700 | DME1701 | |
|---|---|---|---|
| Frequency and Period measurement (FREQ/PERIOD) | For ACV input (all measurement ranges) | Shared with AC voltage measurement (ACV) | |
| For ACI input (all measurement ranges) | Shared with AC current measurement (ACI) | ||
| Continuity test (CONT) | 1000 Ω range (fixed) | 1 % over-range allowed | |
| Diode test | All measurement ranges | No over-range | |
AC voltage measurement (ACV)
| Item | DME1700 | DME1701 | ||||
|---|---|---|---|---|---|---|
| Measurement range | Resolution | Frequency | Accuracy (1 year) | Temperature coefficient/°C | Accuracy (1 year) | Temperature coefficient/°C |
| 100 mV to 750 V | 0.1 μV to 1 mV | 3 Hz to 5 Hz | 1.00+0.03 | 0.100+0.003 | 1.00+0.04 | 0.100+0.003 |
| 5 Hz to 10 Hz | 0.38+0.03 | 0.035+0.003 | 0.35+0.04 | 0.035+0.003 | ||
| 10 Hz to 20 kHz | 0.09+0.03 | 0.005+0.003 | 0.06+0.04 | 0.005+0.003 | ||
| 20 kHz to 50 kHz | 0.15+0.05 | 0.011+0.005 | 0.12+0.05 | 0.011+0.005 | ||
| 50 kHz to 100 kHz | 0.63+0.08 | 0.060+0.008 | 0.60+0.08 | 0.060+0.008 | ||
| 100 kHz to 300 kHz | 4.00+0.50 | 0.200+0.020 | 4.00+0.50 | 0.200+0.020 | ||
AC current measurement (ACI)
| Item | DME1700 | DME1701 | ||||
|---|---|---|---|---|---|---|
| Measurement range | Resolution | Frequency | Accuracy (1 year) | Temperature coefficient/°C | Accuracy (1 year) | Temperature coefficient/°C |
| 10 mA | 10 nA | 3 Hz to 5 Hz | 1.00+0.04 | 0.100+0.006 | 1.00+0.04 | 0.100+0.006 |
| 5 Hz to 10 Hz | 0.30+0.04 | 0.035+0.006 | 0.30+0.04 | 0.035+0.006 | ||
| 10 Hz to 5 kHz | 0.10+0.04 | 0.015+0.006 | 0.10+0.04 | 0.015+0.006 | ||
| 5 kHz to 10 kHz (TYP) | 0.10+0.04 | 0.030+0.006 | 0.10+0.04 | 0.030+0.006 | ||
| 100 mA | 100 nA | 3 Hz to 5 Hz | 1.00+0.04 | 0.100+0.006 | 1.00+0.04 | 0.100+0.006 |
| 5 Hz to 10 Hz | 0.30+0.04 | 0.035+0.006 | 0.30+0.04 | 0.035+0.006 | ||
| 10 Hz to 5 kHz | 0.10+0.04 | 0.015+0.006 | 0.10+0.04 | 0.015+0.006 | ||
| 5 kHz to 10 kHz (TYP) | 0.10+0.04 | 0.030+0.006 | 0.10+0.04 | 0.030+0.006 | ||
| 1 A | 1 μA | 3 Hz to 5 Hz | 1.00+0.06 | 1.000+0.060 | 1.00+0.06 | 1.000+0.060 |
| 5 Hz to 10 Hz | 0.35+0.04 | 0.035+0.006 | 0.35+0.04 | 0.035+0.006 | ||
| 10 Hz to 5 kHz | 0.23+0.04 | 0.015+0.006 | 0.23+0.04 | 0.015+0.006 | ||
| three a (baseball) | 10 μA | 3 Hz to 5 Hz | 1.00+0.06 | 1.000+0.060 | 1.00+0.06 | 1.000+0.060 |
| 5 Hz to 10 Hz | 0.35+0.04 | 0.035+0.006 | 0.35+0.04 | 0.035+0.006 | ||
| 10 Hz to 5 kHz | 0.23+0.04 | 0.015+0.006 | 0.23+0.04 | 0.015+0.006 | ||
AC measurement characteristics
| Item | DME1700 | DME1701 | ||
|---|---|---|---|---|
| Common specifications | AC measurement mode | With anti-aliasing filter AC-coupled digital sampling | ||
| Crest factor (except sine waves) | Crest factor selection | The larger of the following crest factors: • Up to 3:1 of full scale input • Max. 10:1 | ||
| Measurement range selection in auto range | Optimal measurement range for the crest factor up to 10:1 | |||
| Applying the accuracy characteristics | [Crest factor] × [fundamental frequency] ≤ 3 kHz (applied to all crest factors) | |||
| Volt-Hertz product | 8 × 107 V*Hz or below | |||
| Common mode rejection ratio (If LO terminal lead imbalance is 1 kΩ) | >>70 dB | |||
| AC filter | Measuring input signals with frequencies greater than the set AC filter bandwidth | |||
| Effect of AC filter bandwidth on aperture time | • 2 ms at 200 Hz • 20 ms at 20 Hz • 200 ms at 3 Hz | |||
| Input impedance (if parallel connection < 100 pF) | 1.1 MΩ ± 2 % | |||
| Measurement range additional error | If the input is below 50 kHz and is 1 % to 5 % of the measurement range | Adding 0.1 % of the measurement range to the accuracy | ||
| If the input is 50 kHz to 100 kHz | Adding 0.13 % of the measurement range to the accuracy | |||
| AC voltage measurement (ACV) | Over-range | For 100 mV to 100 V ranges | 20 % over-range allowed | |
| For 750 V range | No over-range | |||
| Maximum DC voltage | All ACV measurement ranges | 400 V | ||
| AC current measurement (ACI) | Over-range | For 10 mA to 1 A ranges | 20 % over-range allowed | |
| For 3 A range | No over-range | |||
General Specifications
| Item | DME1700 | DME1701 | |
|---|---|---|---|
| Nominal voltage range (allowable voltage range) | 100 Vac to 240 Vac (90 Vac to 264 Vac) | ||
| Power consumption | 25 VAmax | ||
| Environmental conditions | Operating temperature range / Operating humidity range | 0 °C to 50 °C (32 °F to 122 °F) / 80 %rh or less (no condensation) | |
| Storage temperature range | -40 °C to 70 °C (-40 °F to 158 °F) | ||
| Altitude | Up to 2000 m | ||
| Dimensions / Weight | 212.8(8.38)W × 88.3(3.48)H× 272.3(10.72)D mm(inches) (Main unit only. Protrusions and other parts are not included) / 2.5 kg (5.5 lbs) | ||
| Interface | – | USB, LAN | |
| Safety*1 | Complies with the requirements of the following directive and standards. Low Voltage Directive 2014/35/EU*4 EN 61010-1 (Class I *2, Pollution Degree 2 *3), EN 61010-2-030 Measurement Category II 300 V EN 61010-031 | ||
| Electromagnetic compatibility (EMC)*1 *4 | Complies with the requirements of the following directive and standards. EMC Directive 2014/30/EU, EN 61326-1, EN 61326-2-1, EMI: CISPR 11 (Class A *5, Group I *6), EN 61000-3-2, EN 61000-3-3 EMS: EN 61000-4-2, EN 61000-4-3, EN 61000-4-4, EN 61000-4-5, EN 61000-4-6, EN 61000-4-8, EN 61000-4-11 Applicable under the following conditions: The maximum length of all cabling and wiring connected to the product must be less than 3 m. | ||
| Input protection | UT HI terminal | 1000 Vpk | |
| SENSE HI terminal | 350 Vpk (referenced to LO terminal) | ||
| SENSE LO terminal | |||
| Insulation resistance (Between each terminal and chassis) | 500 Vpk, >>10 GΩ | ||
| Capacity (Between each terminal and chassis) | < 300 pF | ||
*1 Does not apply to specially ordered or modified products.
*2 This is a Class I instrument. Be sure to ground this product’s protective conductor terminal. The safety of this product is guaranteed only when the product is properly grounded.
*3 Pollution is addition of foreign matter (solid, liquid or gaseous) that may produce a reduction of dielectric strength or surface resistivity.
Pollution Degree 2 assumes that only non-conductive pollution will occur except for an occasional temporary conductivity caused by condensation.
*4 Only for models with CE marking / UKCA marking on their body.
*5 This is a Class A instrument. This product is intended for use in an industrial environment.
This product may cause interference if used in residential areas. Such use must be avoided unless the user takes special measures to reduce electromagnetic emissions to prevent interference to the reception of radio and television broadcasts.
*6 This product belongs to Group 1 products. This product does not generate and/or use intentionally radio-frequency energy, in the form of electromagnetic radiation, inductive and/or capacitive coupling, for the treatment of material or inspection/analysis purpose.
Options
10-channel multi-point scanner card
For DME1701
20-channel multi-point scanner card
For DME1701
10-channel thermocouple multi-point scanner card
For DME1701
10-channel scanner/10-channel relay card
For DME1701
RS232C card
For DME1701
GPIB card
For DME1701
Kelvin probe (for 4-wire resistance measurement)
4-wire test lead
Catalog
Digital Multimeter – DME1700 Series
https://global.kikusui.co.jp/download/cat_dme1700/