When unlocking the interlock on TOS5200 via the SIGNAL I/O connector, why does the interlock protection occur?

Interlock on/off involves opening or short-circuiting pins 1 and 13 of the SIGNAL I/O connector. This operation causes a logical change that triggers the protection state. It can be cleared by inputting a STOP signal during the short circuit.