What is the combination of using PCR-LE series to output 220 V and 50 A at 50 Hz in a three-phase configuration?

For three-phase output, you need three PCR9000LE units and optional 3P05-PCR-LE (three-phase output driver).
To obtain a line voltage of 220 V, you can set to output 127 V in the L range. Alternatively, you can use the H range, but please note that the output current capacity will be lower compared to the H range.
If the power factor is up to 0.85, you can obtain an output of 70.8 A (calculated as 9000 / 127).
However, it is important to note that PCR6000LE can only provide 47.2 A, which falls short of the required 50 A.