How do I use the sequence function of PCR-LE series?

Here is an overview of the sequence function of PCR-LE series:
The sequence can consist of up to 600 steps.
When the sequence is executed, the steps are executed in order, starting from the specified starting step.
Each step allows the configuration of 22 items, such as AC voltage, frequency, and DC voltage:
To access the step editing screens, press the SEQ (SHIFT + SIM) key > the EDIT (F5) key.
There are six types of step editing screens, and you can navigate through them using the F6 key to move through the items, and select the item to be set. Press the ESC key to exit the pages and return to the previous screen for other settings.
Before executing the sequence, set the conditions as follows:
Press the SEQ (SHIFT + SIM) key > the COND (F4) key to set the start and end steps.
Press the SEQ (SHIFT + SIM) key > the LOOP (F3) key to specify the number of repetitions.
To execute the sequence, follow these steps:
Press the SEQ (SHIFT + SIM) key > the RUN (F1) key to start the sequence.
For more details, read the user’s manual – chapter of Using the Sequence Function in Advanced.