TOS5400_Withstanding Voltage Test Section-Measurement function

Digital voltmeter

ItemTOS5401TOS5402TOS5403
Measurement rangeAC/DC: 0.000kV to 6.500kVAC: 0.000kV to 6.500kVDC: 0.000kV to 6.500kV
Display□. □□□ kV
AccuracyV < 500V: ±(1.5% of reading + 10V)
V ≥ 500V: ± 1.5% of reading
Response*1ACW: True rms/ Mean-value response rms display can be switched.
DCW and IR: Mean-value
Mean-value
Hold featureHolds the measured voltage value at the end of the test while the PASS/FAIL judgment result is displayed.

*1 The true rms value and the average response rms value, require a response time of at least 50 ms in order to meet the measurement accuracy.

Digital ammeter

ItemTOS5401TOS5402TOS5403
Measurement rangeAC: 0.00mA to 110mAAC: 0.00mA to 110mADC: 0.00mA to 11mA
DC: 0.00mA to 11mA
Display (i = measured current)i < 10uA□. □μA
10uA ≦ i < 100uA□□. □μA
100uA ≦ i < 1mA□□□. □μA
1mA ≦ i < 10mA□. □□□mA
10mA ≦ i < 100mA□□. □□mA
100mA ≦ i□□□. □mA
Accuracy*1 (i = measured current)1.00mA ≦ i±(1.5% of reading)
i < 1.00mA±(1.5% of reading + 20 μA)
Response*2ACW: True rms/ Mean-value response rms display can be switched.Mean-value
Hold featureThe current measurement after a test is finished is held while the pass judgment is displayed.

*1 When the humidity is 70% or higher, add 50μA. During AC voltage tests, current also flows in the stray capacitance of items such as the test leads and tools. For details on stray capacitance, see “Stray Capacitance of AC Withstanding Voltage Tests”.
*2 The true rms value and the average response rms value, require a response time of at least 50 ms in order to meet the measurement accuracy.