TOS5400_Withstanding Voltage Test Section-Judgment functionÄb0

ItemTOS5401TOS5402TOS5403
Current judgment*1FunctionThe output is shut off when a judgment is made. Beep volume level can be set for PASS and FAIL separately.
UPPER FAILJudgment methodUpper limit (Upper) is detected.
DisplayShown on the display.
Beep soundOn
SIGNAL I/OThe U FAIL signal is generated continuously until a STOP signal is received.
LOWER FAILJudgment methodLower limit (Lower) is detected. Voltage rise time (Rise Time)Voltage fall time (Fall Time)
DisplayShown on the display.
Beep soundOn
SIGNAL I/OThe L-FAIL signal is generated continuously until a STOP signal is received.
PASSJudgment methodPASS judgment is made if Upper-FAIL or Lower-FAIL has not occurred when the test time elapses.
DisplayShown on the display.
Beep soundON (fixed at 0.2 s regardless of Pass Hold setting)
SIGNAL I/OThe PASS signal is generated for the length of time specified by the Pass Hold setting. If Pass Hold is set to Infinity, the PASS signal is generated continuously until a STOP signal is received.
Upper limit setting rangeAC: 0.01mA to 110.00mA DC: 0.01mA to 11.00mADC: 0.01mA to 11.00mA
Lower limit setting rangeAC: 0.00mA to 109.99mA, or OFF DC: 0.00mA to 10.99mA, or OFFDC: 0.00mA to 10.99mA, or OFF
Judgment accuracy*21.00mA ≦ i±(1.5% of setting)
i < 1.00mA±(1.5% of setting + 20 μA)
Current detection methodCalculates the current’s true rms value or mean-value and compares this value with the reference value
CalibrationCalibrated with the rms of a sine wave using a pure resistive load
Discharge detection functionFunctionSeparately from the Upper judgment, it is possible to choose between low-pass filter (LPF) and high-pass filter (HPF) to detect a discharge at the EUT.
LPFThe judgment criterion is set to a value greater than the set Upper value. It can be switched between three levels: Fast, Medium, Slow. It is generally for normal withstanding voltage test.
HPFThe judgment criterion is set to a fixed value independently of the set Upper value. It can be switched between three levels: High, Medium, Low. It is mainly for detecting small discharges.

*1 In the following conditions, charging current may activate a FAIL judgment or trigger the protection function. EUT capacitance: 1μF; Rise Time: 1 second or less; Judge Delay: 1.2 seconds or less.
*2 When the humidity is 70% or higher, add 50μA. During AC voltage tests, current also flows in the stray capacitance of items such as the test leads and tools. For details on stray capacitance, see “Stray Capacitance of AC Withstanding Voltage Tests”.