ModelPLZ164WPLZ334WPLZ1004WPLZ164WAPLZ664WA
Overvoltage protection (OVP)Turns off the load at 110 % of the rated voltage
Overcurrent protection (OCP)0.03 A to 36.3 A0.06 A to 72.6 A0.2 A to 220 A0.03 A to 36.3 A0.13 A to 145.2 A
Or 110 % of the maximum current of each range
Overpower protection (OPP)0.1 W to 181.5 W0.3 W to 363 W1 W to 1100 W0.1 W to 181.5 W0.6 W to 726 W
Or 110 % of the maximum power of each range Load off or limit selectable
Overheat protection (OHP)Turns off the load when the heat sink temperature reaches 95 °C
Undervoltage protection (UVP)Turns off the load when detected.
Reverse connection protection (REV)By diode and fuse. Turns off the load when an alarm occurs.